VIDEO TRANSCRIPT

ZX2 Product Overview

This video is an overview of the ZX2 loop modulation system.  Here we will describe the major assemblies and their functions. Please note that, for illustration purposes, not all pneumatic hoses and electrical cables are shown.   

The ZX2 retrofits onto several commercially available GC systems, including the Agilent models 7890 A or B and 8890 gas chromatographs. 

A floor chiller cools a stream of room temperature, dry nitrogen gas to about -90 degrees Celsius, suitable for thermal modulation of compounds in the C9 plus range.  The user supplies the lab bench on which the GC is installed.  Easy access to the back of the GC is required. 

ZX2 Thermal Modulator: floor chiller (no cryogenic consumables)

A hot jet valve mounted on top of the GC, supplies gas pulses to a hot jet assembly, located inside the GC oven.  A cold jet transfer line supplies refrigerated gas to a cold jet assembly, also located inside the GC oven. 

ZX2 Thermal Modulator: cold jet transfer line, hot jet valve

This is a side view of the cold jet transfer line, inside of which the cold jet gas stream is refrigerated. A manually operated cold jet flow controller sets the cold jet flow rate – typically 10-15 liters per minute.

ZX2 Thermal Modulator: cold jet transfer line, cold jet flow controller

The floor chiller rests behind and below the bench top, as shown here.

ZX2 Thermal Modulator: floor chiller (no cryogenic consumables)

A cold jet valve switches the cold gas supply on and off. Typically, the cold jet valve is turned off for the first minute of a run, turned on for the duration of a run, and left on between runs. The kit includes a utility valve, which may be used to decrease the cold jet flow rate late in a run, a technique used for analysis of heavier organic species.  (Contact Zoex for more details.)

ZX2 Thermal Modulator: Cold Jet Valve

A tee connector upstream of the utility valve feeds room temperature dry nitrogen gas to the hot jet valve, located on top of the GC.

ZX2 Thermal Modulator: Tee to hot jet valve

A pressure regulator supplies 40 psi of house nitrogen to the valve pack.  

Zx2 Thermal Modulator: Pressure Regulator

The nitrogen inlet is a 1/4-inch Swagelok fitting, which receives house nitrogen at a feed pressure of 100 psi.

Zx2 Thermal Modulator: Nitrogen inlet

A programmable logic controller, or “PLC”, draws power from a DC power supply included in the kit. The PLC issues on-off signals to the hot jet valve, which effects the thermal modulation process. 

Zx2 Thermal Modulator: Nitrogen Inlet

The hot jet valve draws power from a valve control port on the GC motherboard. 

Zx2 Thermal Modulator: valve control port

A touchscreen human-machine-interface, or HMI, controls the PLC.  With the HMI touchscreen, the user sets the modulation period (i.e., the time between hot jet firings) and the duration of hot jet pulses, which remobilize trapped substances. A modulation period of 6 seconds and a pulse duration of 0.375 seconds are typical.

ZX2 Thermal Modulator: HMI for Modulation Control

This is a view of the modulator assembly, mounted in the GC oven. Analytical columns are not shown.

The modulator tube consists of a length of column, two portions of which pass through intersecting hot and cold gas jets. These two column portions form the so-called “loop”.

ZX2 Thermal Modulator: modulator tube: "loop"

The cold jet, which is vacuum insulated, protrudes downward from the oven roof, passes through the head clamp, and then terminates about 2 mm above the loop “apex”, or extremum. There, the cold and hot jets periodically trap and remobilize analytes eluting from the primary column of the GCxGC column set.

ZX2 Thermal Modulator: cold jet, hot jet

 The modulator tube is wound into a “column holder” which is part of the column set, and attaches to the head clamp.

ZX2 Thermal Modulator: column holder

An insulated canister houses a hot jet body, a cartridge heater, and a platinum resistance temperature sensor. 

ZX2 Thermal Modulator: insulated cannister

Hot jet temperature is controlled by a GC auxiliary thermal zone via a heater/sensor cable.  A convoluted steel gas line feeds dry nitrogen gas to the hot jet, which pulses once at the start of each modulation cycle.

ZX2 Thermal Modulator: heater/sensor cable, hot jet gas line

Repetitive cooling and heating of the loop extremum results in a GCxGC chromatogram. 

To learn more, contact Zoex: info@zoex.com

ZOEX
To get a quote or learn more about our products and services, contact us. One of our representatives will assist you.
Contact Information
LocationIcon

12221 Fuqua Street, Suite A,
Houston, TX

EmailIcon

info@zoex.com

PhoneIcon

+1 281 484 2828 (office)

MobileIcon

+1 866 904 2942 (toll free)

FaxIcon

Fax: +1 281 484 0808

ZOEX
To get a quote or learn more about our products and services, contact us. One of our representatives will assist you.
Contact Information
LocationIcon

12221 Fuqua Street,
Suite A,
Houston, TX

EmailIcon

info@zoex.com

PhoneIcon

+1 281 484 2828
(office)

MobileIcon

+1 866 904 2942
(toll free)

FaxIcon

Fax: +1 281 484 0808